As good as new Hach FP360 sc PAH/oil fluorescence probe, LXV441.99.11101 , 0-500 ppb, stainless steel, 10 m with flow cell on mounting plate, LZY669
Fluorescence probe for measuring PAH/oil in water. Stainless steel housing, 10 m cable, without cleaning nozzle.
Even the smallest traces of oil affect water quality. The probe continuously monitors water, process and industrial water for mineral oil contamination down to trace levels. The highly sensitive UV fluorometer is immersed directly into the medium.
Fluctuations in the intensity of the flash lamp are compensated for to ensure reliable and long-term stable hydrocarbon monitoring. Daylight influences are automatically eliminated.
The probe can be combined with other sensors on the SC controllers.
The right technology at the right price
Minimal maintenance effort
Less testing in the laboratory
High sensitivity and selectivity
High sensitivity and selectivity
The FP 360 sc can detect polycyclic aromatic hydrocarbons
hydrocarbons (PAHs) from 1 ppb up to 5000 ppb
(μg/L) can be detected and measured. This corresponds approximately
a mineral oil concentration of 0.1 to 150 ppm (mg/L).
In addition, the FP 360 sc detection method is
insensitive to interference from turbid water
or natural organic and biological substances that are
online light scattering, UV absorption and VIS fluorescence instruments.
fluorescence instruments.
Measuring method
UV fluorescence method for polycyclic aromatic
aromatic hydrocarbons (PAH)
Light source
Miniature xenon flash lamp with interference filter
Detector
UV photodiode with interference filter;
Compensation of fluctuations in
daylight and flash lamp intensity
Excitation wavelength
254 nm
Measurement wavelength
360 nm
Measuring range
Low measuring range:
0-50 ppb (μg/L) and 0-500 ppb (μg/L) (PAH)** 0.1-
1.5 ppm (mg/L) and 0.1-15 ppm (mg/L) (oil)**
High measuring range:
0-500 ppb (μg/L) and 0-5,000 ppb (μg/L) (PAH)**
0.1-15 ppm (mg/L) and 0.1-150 ppm (mg/L) (oil)**
Resolution
0.1 ppb (μg/L) (PAH) in the lowest measuring range
Limit of detection (LOD) is 1 ppb (PAH)
Reproducibility
2.5% of the measured value at constant temperature
Response time
10 s (T90)
Calibration
Factory calibration with UV fluorescence standard or
Process calibration with results of a random sample analysis.
Sample temperature
33.8 to 104°F or 1 to 40°C
Pressure range
Max. 30 bar or 435 psia (measuring probe)
Housing
Stainless steel 316Ti (1.4571) or titanium
Dimensions
2.68“ x 12.05” or 68 x 306 mm
(D x H; without connector and suspension bolt)
Weight
Stainless steel: 6.2 lbs or 2.8 kg
Translated with DeepL.com (free version)