Memmert WNE22 Waterbath
temperature
Setting temperature range +10 to +95 ° C and cooking level
Working temperature range in ° C At least 5 above room temperature to + 95 ° C with an additional cooking setting
Display resolution / setting accuracy 0.1 ° C to 99.9 ° C / 1 ° C from 100 ° C
Standard component control
Controller digital display of all default values: temperature and alarm values (resolution 0.1 ° C), as well as time values
Timer digital timer from 1 min. To 999 hours for: ON (continuous operation) WAIT (delayed on) HOLD or HOLD target temperature-dependent with "holding time guarantee"
Calibration on the controller
safety
Temperature sensor 2 Pt100 sensors DIN class A in 4-wire measurement with mutual monitoring and function takeover at exactly the same temperature value
Temperature monitoring of mechanical temperature limiter TB protection class 1 to switch off the heating approx. 30 ° C above the maximum bath temperature
Temperature monitoring of separately working electronic temperature safety monitors TWW protection class 3.1 or temperature limiters TWB protection class 2 can be set by the customer in the set-up
Temperature monitoring in the event of excess temperature caused by a fault, the heating is switched off at a fixed interval of approx. 10 ° C above the setpoint
Self-diagnosis system PID microprocessor control with fuzzy support and self-diagnosis system for quick troubleshooting
Alarm optical and acoustic alarm at the end of the program and for input acknowledgment as well as if the fill level is too low, the heating is switched off automatically
Heating concept
Heating baths, corrosion-protected large-area heating on 3 sides
Stainless steel interior
Dimensions W (A) x H (C) x D (B): 350 x 220 x 290 mm
Interior easy to clean, stiffened by deep-drawn ribbing, W.-St. 1.4301 (ASTM 304), laser welded
Volume 22 l
Fill level at least 177 mm
Fill level max. 200 mm
Structure stainless steel case
Dimensions W (D) x H (G) x D (E): 578 x 405 x 436 mm
Electrical data
Voltage 230 V, 50/60 Hz
Power consumption approx. 2000 W (in heating function)
Environmental conditions
Installation The distance from the wall on all sides must be at least 80 mm. There must be a clearance of at least 750 mm upwards from the upper edge of the bath to the next delimiting area.
Ambient temperature +5 ºC to +40 ºC
Humidity rh max. 80% non-condensing
Overvoltage category II
Pollution degree 2
Condition: very well preserved.