New, factory-sealed Huber Unichiller 100Tw-H with extended temperature range from -20° to 130°, 10kW cooling capacity at 15°
Temperature range -20...130 °C
Temperature stability ±0.2 K
Controller Pilot ONE
Safety class III / FL
Protection class IP20
Dimensions (W x D x H) 600 x 600 x 1450 mm
Weight, net 235 kg
Heating capacity 2 kW
Cooling capacity
15 10
0 10
-10 6.5
-20 3
°C
kW
Refrigeration machine water-cooled, CFC- and HCFC-free
Refrigerant 1st stage R-449A (A1, H280)
Refrigerant quantity 1st stage 1.75 kg
External sensor connection Pt100
Digital interface Ethernet, USB (host and device), RS232
Permissible ambient temperature 5...40 °C
Sound pressure level 65 dB(A)
Minimum filling volume 6.5 l
Expansion volume 5.5 l
Cooling water connection G3/4 AG
Cooling water consumption with water 15°C, flow -20°C 276 l/h
Cooling water consumption with water 15°C, flow -10°C 375 l/h
Cooling water consumption with water 15°C, flow 0°C 510 l/h
Cooling water consumption at water 15°C, flow 15°C 510 l/h
Cooling water differential pressure min. 1 bar
Cooling water pressure max. 6 bar
Mains connection 400V 3~ 50Hz
Current consumption max. 15 A
Fuse protection 3x20A
Translated with DeepL.com (free version)