Photometer Anthos HT-3
Anthos HT-3, the fully automatic temperature-controlled microplate reader with versatile,
sophisticated internal software on a graphics display. Comes with Stingray software from DAZDAQ for evaluation and dustcover.
Methods of measurement:
Mono-and bichromatic endpoint measurement
Kinetics: Up to 40 mono-, bichromatic measurements without external PC
Special two-point kinetics scanning measurement: On screen or remote control
Results aud display, printer or serial interface
Fast methods of measurement:
Mono-, bi-and multichromatic measurement with the issue of OD values on display,
Printer or serial interface, kinetic reaction process on display.
Flexible features for disk usage:
Arbitrary Blank validity, and fill direction replikation programming.
Up to four controls and 12 standards can be positioned on the plate.
Procedures for the evaluation:
All evaluation can measurements, results of kinetic data reduction
and are based relational values. Before printing, the results shown on the display and outliers are eliminated.
Powerful formula entry:
The results of measurements of all special items on the plate, and Mathematical Constants
Operators can in formulas for the calculation of threshold values, standards (eg pre-set
be standard curves, single point) and relational values (eg B/B0 used).
Qualitative evaluation:
Up to 5 areas
Car or% division between two thresholds
Individual name environment for areas
Quantitative analysis:
Point to Point
Linear regression
Cubic Spline
4-Parameter Fit
All evaluation method can be combined
Quantitative results can be qualitatively of classified.
Kinetic data reduction:
ΔOD, middle slope,
Maximum gradient
ΔOD max. Slope,
Δ t max slope
Δ t relative,
Δ t absolute.
Special opportunities:
Plate centering position any type of micro-plates under the precise measurement beam Programmable shaking
Functions before the measurement:
Observation of selected items before the measurement, shaking, enter a dilution factor
Automatic self test on power:
Light source, detector and internal software
Identification plate manually or via bar code (optional from the factory)
Data storage:
30-50 tests (flexible memory management), 10 plates (measured values or kinetic data reduction) for later analysis
Interface:
Graphic Display: 16x42 characters (128x256 pixel)
Keyboard with function keys
Parallel printer interface (IBM or EPSON compatible 9 - and 24 - Dot Matrix Printers) Serial interface to control all the measurement functions, transmission: measurements, evaluation results of measured plates, remote start of the measurement, evaluation and data transfer.
Key Applications:
Routine laboratories, blood banking, research, computer based systems and many others.
TÜV GS tested according to E DIN VDE 0411 part 100/08.86 and DIN VDE 0750 Part 1
Technical data:
Photometric Measurement Method: Continuous photo-optical sensing to 8 measuring and one reference channel. Self-calibration before each measurement
Light source: Tungsten halogen lamps Regulated
Measuring speed:
Monochromatic: 5 seconds
Eichromatisch: 7 seconds
Tempering:
Temperature adjustable to 30 ° / 37 °
At 22 ° (+ / -1 °) room temperature is the temperature of the measuring chamber + / -0.1 ° at 30 ° and + / -0.2 ° at 37 ° stable.
Power supply:
Current consumption: 140 VA (with all options)
90-130 Volt / 180 - 260 Volts
Frequency: 47-63 Hz
(Self)
Interface:
Serial interface: 9-pin, 300 to 19200 baud
Parallel printer interface: 25-pin
Dimensions:
34cm x 23cm x 44cm (W x H x L)
Weight: 18.5 kg net
Condition: Used