Technical data:
Ambient temperature:
Storage - 25 ° C ... + 65 ° C
Operating 0 ° C ... + 55 ° C
Permissible relative
Humidity
Year: <75%
30 days / year: 95%
Other days: 85%
Ranges and
Resolutions:
[US / cm]
[MS / cm]
0.000 ... 1,999
(Only for cell. = 0.01 cm-1)
0.00 ... 19,99
(Only for cell. = 0.01 cm-1
and cell const. = 0.1 cm-1)
0.0 ... 199,9
0 ... 1999
0.00 ... 19,99
0.0 ... 199,9
0 ... 500
spec. Resistance
[Milliohms * cm]
0.000 ... 1,999
0.00 ... 19,99
0.0 ... 199,9
0 ... 1999
SAL 0.0 ... 70.0 according to the IOT table
TDS [mg / l] 0 ... 1999
Factor set between
0.40 ... 1,00
T [° C] - 5.0 ... + 105,0
T [° F] + 23.0 ... + 221.0
Accuracy (± 1 digit):
68
No compensation:
Accuracy ± 0.5%
Nonlinear compensation
Accuracy Sample
± 0.5% 0 ° C ... 35 ° C
to EN 27 888;
± 0.5% 35 ° C ... 50 ° C
Advanced
nLF function after
WTW measurements
Linear Compensation:
Accuracy Sample
± 0.5% 10 ° C ... 75 ° C
(Refers to the percentage of accuracy
in each case to the measured value!)
SAL range 0.0 ... 42,0
Accuracy Sample
± 0.1 at 5 ° C ... 25 ° C
± 0.2 at 25 ° C ... 30 ° C
TDS [mg / l] 1
T [° C] NTC 30:
± 0.1
PT 1000:
Operating accuracy
± 0.5 0 ° C ... 15 ° C
± 0.1 15 ° C ... 35 ° C
± 1 35 ° C ... 55 ° C
T [° F] NTC 30:
Accuracy ± 0.2
PT 1000:
Operating accuracy
± 0.9 at 32 ° F ... 59 ° F
± 0.2 at 59 ° F ... 95 ° F
± 1.8 at 95 ° F ... 131 ° F
Setting the cell constant:
C [cm-1] 0.010
0.090 ... 0,110
0.250 ... 2,500
Calibrate the cell constant:
C [cm-1] 0.450 ... 0,500
0.800 ... 1,200
Selectable reference temperature:
TREF [° C] 20
25
Temperature input:
Manually [° C] - 5 ... +100
Dimensions and weight:
Length [mm] 250
Width [mm] 300
Height [mm] 70
Weight [kg] 1.3 (without AC adapter)
Analog output:
Automatic switch-over when the cable,
AK323.
Signal range
0 ... 1.999 V for the range
0 ... 1999 digits
Accuracy ± 0.5% of reading
± 0.1 (% saturation)
or
± 0.5% of reading ± 0.01 mg / l
Internal resistance
<5 Ohm
(Current limited to max. 0.2 mA output current)
Serial interface:
Automatic switch-over when the cable
AK340 / B, AK325 / S
Type RS232, data output
Baud rate set to 1200, 2400, 4800, 9600
Data bits 8
Stop bit 1
No parity (None)
RTS / CTS + Xon / Xoff
Cable length max. 15 m
Power Supply:
Batteries 4 x 1.5 V
Alkaline manganese batteries type AA
Running time approx 3000 hours
Network for the power supply unit is:
Connection max. Overvoltage category II
Power supply with Euro plug:
FRIWO FW3288, 11.8134
Friwo Part No. 1816492
Input: AC 230V / 50 Hz / 23 VA
Output: 6 V = / 1.8A / 10.8 VA
AC adapter with U.S. plug:
FRIWO FW3288, 11.8451
Friwo Part No. 1816493
Input: 120V ~ / 60 Hz / 21.5 VA
Output: 6 V = / 1,8 A
AC adapter with UK plug:
FRIWO FW3288, 11.8453
Friwo Part No. 1770896
Input: AC 230V / 50 Hz / 23 VA
Output: 6 V = / 1,8 A
Guidelines and standards:
EMC directive 89/336/EEC
EN 61326-1:1997
EN 61000-3-2 A14: 2000
EN 61000-3-3:1995
FCC Class A
Equipment Safety Council Directive 73/23/EEC
Class 3, EN 61010-1 A2: 1995
Climate Class 2, VDI/VDE3540
Certified:
TÜV GS, UL / CUL, CE